Shimpo DFS-series Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Werkzeug Shimpo DFS-series herunter. SHIMPO DFS-series User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 20
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
Congratulations on your purchase of a Shimpo DFS (Digital
Force Gauge). We trust you will enjoy many years of
professional results from your DFS for force measurement
applications.
Please read the entire instruction manual thoroughly before
initial set-up and operation; the information contained
herein will aid you in operating your Shimpo DFS safely and
with excellent results.
If you have any questions regarding our product(s), call your
local Shimpo representative or contact Shimpo directly for
assistance.
DFS-Digital Force Gauge
Instruction ManualInstruction Manual
Instruction ManualInstruction Manual
Instruction Manual
Contents
Inspection / Standard Accessories 2
Features and Benefits 2
Important Safety Instructions 2
Set-Up 3
Power 3
Display and Keypad 4
Factory Settings 5
Operation 5-8
How to Tare and Zero the Display 5
Changing the Measuring Mode 5
Changing Units of Measure 6
Setting High and Low Limits 6
Changing Display Update Time 6
Changing Memory Modes 7
Recalling Statistics 8
Setting of Filter 8
Communication with External Devices 9-16
Output Selection 9-13
RS232 10-12
Digimatic 13
Analog 13
Comparator 13
External Control Input 14-15
Overload Output 16
Parameter Settings 17
Low Battery Indication 18
Auto Power Off 18
Calibration 18-19
Troubleshooting 19
Dimensions and Specifications 20
Warranty 20
Seitenansicht 0
1 2 3 4 5 6 ... 19 20

Inhaltsverzeichnis

Seite 1 - Instruction Manual

Congratulations on your purchase of a Shimpo DFS (DigitalForce Gauge). We trust you will enjoy many years ofprofessional results from your DFS for fo

Seite 2 - Features and Benefits

10Communication With External Devices (continued)RS232Depending upon the type of data collection device that is being used with the DFS, it may be nec

Seite 3

11External Control CommandsThe chart below lists the ASCII Capital Characters andtheir functions for controlling the DFS from a computeror other exter

Seite 4 - Display & Keypad

12Communication With External Devices (continued)External Control Command Descriptions (continued)! APF ____________________ Auto-Power-OffAPF # CR

Seite 5 - Operation

13Communication With External Devices (continued)DigimaticDigimatic output is used for Mitutoyo data collectors and printers (refer to RS232 memory da

Seite 6 - Operation (continued)

14Communication With External Devices (continued)External Control InputNOTE: To activate the comparator cable, either “RS232” or “nonE” needs to be se

Seite 7

15Communication With External Devices (continued)External TareTo tare, short pins 5 and 10 for at least 25mS.External HoldChoose edge or level trigger

Seite 8

16Communication With External Devices (continued)Overload OutputThe DFS possesses an overload output feature (DFS-CTRLCABLE) to protect the load cell

Seite 9 - Output Selection

17Parameter SettingsThe DFS will default to certain parameters when power is turned OFF and ON. To access the parameter settings:1. Press the PEAK bu

Seite 10

18Low Battery IndicationThe batteries should be changed when “LO BAT” appears on the display.NOTE: Parameters and data in memory can be saved if the A

Seite 11

19TroubleshootingThe following are general checkpoints; please call your local Shimpo representative or contact Shimpo Instruments directly forfurther

Seite 12

2Important Safety InstructionsInspection/Standard AccessoriesIf any damage is apparent, do not unpack the DFS. Notifythe shipping carrier immediately

Seite 13

20WarrantyLIMITED EXPRESS WARRANTY: Shimpo Instruments warrants, to the original purchaser of new products only, that this product shall be free from

Seite 14 - External Control Input

3Set-UpPowerEvery DFS unit is shipped from the factory without the batteries (provided) installed. To install the batteries:1. Remove the battery pla

Seite 15

4Display & Keypad 1. Sensor Shaft 2. Low Battery Indicator 3. Peak Indicator (NOTE: Refer to “Changing the Measuring Mode” section for explanat

Seite 16 - Overload Output

5Factory SettingsEach time the power is turned OFF and ON there are some parameters that the DFS will default to. Consult the sectionon “Parameter Se

Seite 17 - Parameter Settings

64. Press MEMO SET button to store the measuring mode and exit.NOTE: By keeping the true peak mode “ON”, the DFS will always display the peak samples

Seite 18 - Calibration

7Operation (continued)Changing Display Update TimeThe display update time works with the sampling rate (1000 hz) to provide an average reading. All s

Seite 19 - Calibration (continued)

8Operation (continued)Deleting Data (Single and Continuous Memory Mode)While recalling data (see above) it is possible to delete one data reading at a

Seite 20 - Dimensions and Specifications

9Communication with External DevicesWhen operated with the proper cable, the DFS offers a variety of output and input abilities for bi-directional com

Kommentare zu diesen Handbüchern

Keine Kommentare